<Time>, <Frequency>
Frequency drift:
Coordinates of the frequency drift trace as a list of comma sepa-
rated values, beginning at the left border of the display.
<Time1>,<Frequency1>,<Time2>,<Frequency2>,...
Example:
TRAC? TRACE1
Queries the data of trace 1.
Usage:
Query only
Manual operation:
See
See
8.11 Configure Numerical Result Displays
The following commands are necessary to configure the numerical phase noise result
displays.
Configuring Residual Noise Measurements
..........................................................126
Reading Out Residual Noise Results
....................................................................129
Configuring Spot Noise Measurements
................................................................ 131
.....................................................................................134
.............................................................................135
Reading Out the Sweep Result List
...................................................................... 136
8.11.1 Configuring Residual Noise Measurements
...................................................................................126
CALCulate<n>:EVALuation[:STATe]
................................................................................ 127
................................................................................... 127
CALCulate<n>:EVALuation:USER<range>:STARt
.............................................................127
CALCulate<n>:EVALuation:USER<range>:STOP
............................................................. 128
CALCulate<n>:EVALuation:USER<range>:TRACe
............................................................128
CALCulate<n>:EVALuation:STARt
<OffsetFrequency>
This command defines the start point of the residual noise integration range.
Before you can use the command, you have to turn on the measurement range inte-
gration with
CALCulate<n>:EVALuation[:STATe]
Suffix:
<n>
.
Parameters:
<OffsetFrequency>
The minimum offset is 1
Hz. The maximum offset depends on
the hardware you are using.
*RST:
1 kHz
Example:
CALC:EVAL:STAR 1 kHz
Defines an start point of 1 kHz for the residual noise range.