Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
136
User Manual 1173.9286.02 ─ 15
FETCh:PNOise:MEASured:FREQuency?
This command queries the carrier frequency that has been actually measured.
The measured frequency is shown in the channel bar.
Return values:
<Frequency>
Frequency in Hz.
Example:
FETC:PNO:MEAS:FREQ?
Queries the measured frequency.
Usage:
Query only
FETCh:PNOise:MEASured:LEVel?
This command queries the level of the DUT that has been actually measured.
The measured level is shown in the channel bar.
Return values:
<Level>
Level in dBm.
Example:
FETC:PNO:MEAS:LEV?
Queries the measured level.
Usage:
Query only
8.11.6 Reading Out the Sweep Result List
.........................................................................................136
...................................................................................... 137
...................................................................................... 137
..................................................................................... 137
.....................................................................................138
......................................................................................138
.......................................................................................138
FETCh:PNOise:SWEep:AVG?
This command queries the number of measurements that have been performed in
each half decade.
Return values:
<Measurements>
Number of measurements as displayed in the Sweep Result
List.
The command returns one value for each half decade as a
comma separated list.
Example:
FETC:PNO:SWE:AVG?
would return, e.g.:
10,10,10,10,10
Usage:
Query only
Configure Numerical Result Displays