Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
113
User Manual 1173.9286.02 ─ 15
CALCulate<n>:LIMit<k>:UPPer:STATe
<State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
8.9.3 Creating and Editing Standard Limit Lines
.................................................................................. 113
CALCulate<n>:LIMit<k>:CONTrol[:DATA]
.........................................................................113
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
...........................................................................114
CALCulate<n>:LIMit<k>:LOWer[:DATA]
........................................................................... 114
CALCulate<n>:LIMit<k>:LOWer:MODE
............................................................................114
CALCulate<n>:LIMit<k>:LOWer:SHIFt
............................................................................. 115
CALCulate<n>:LIMit<k>:UPPer:MODE
.............................................................................115
CALCulate<n>:LIMit<k>:UPPer[:DATA]
............................................................................116
CALCulate<n>:LIMit<k>:UPPer:SHIFt
.............................................................................. 116
CALCulate<n>:LIMit<k>:COMMent
<Comment>
This command defines a comment for a limit line.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Comment>
String containing the description of the limit line.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol[:DATA]
<LimitLinePoints>
This command defines the horizontal definition points of a limit line.
Suffix:
<n>
.
irrelevant
<k>
Using Limit Lines