Analysis
R&S
®
FSW-K40
63
User Manual 1173.9286.02 ─ 15
Edit Limit Line
The "Edit Limit Line" dialog box contains functionality to describe the shape of a limit
line.
Because limit lines have to meet certain conditions for phase noise measurements, the
availability of parameters is limited.
Name ← Edit Limit Line
Defines the name of a limit line.
Remote command:
Comment ← Edit Limit Line
Defines a comment for the limit line.
A comment is not mandatory.
Remote command:
X-Axis ← Edit Limit Line
Defines the characteristics of the horizontal axis.
The characteristics consist of the unit, the scaling and the type of values.
In the Phase Noise application, the unit for the horizontal axis is always Hz. The scal-
ing can either be logarithmic or linear
"Unit"
In the Phase Noise application, the unit is always Hz.
"Scaling"
In the Phase Noise application, the scaling of the horizontal axis is
always logarithmic.
"Type of Val-
ues"
The type of values can be absolute values or relative to the nominal
frequency.
Using Limit Lines