Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
115
User Manual 1173.9286.02 ─ 15
Suffix:
<n>
.
<k>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
CALCulate<n>:LIMit<k>:LOWer:SHIFt
<Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
<k>
Parameters:
<Distance>
Defines the distance that the limit line moves.
Manual operation:
See
CALCulate<n>:LIMit<k>:UPPer:MODE
<Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
<k>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
Using Limit Lines