Configuration
R&S
®
FSW-K40
47
User Manual 1173.9286.02 ─ 15
Functions to configure the sweep described elsewhere:
●
●
................................................................................... 47
...........................................................................................47
................................................................................................47
................................................................................................. 47
Continuous Sweep /RUN CONT
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
While the measurement is running, the "Continuous Sweep" softkey and the RUN
CONT key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. Furthermore, the RUN CONT key controls the Sequencer, not indi-
vidual sweeps. RUN CONT starts the Sequencer in continuous mode.
For details on the Sequencer, see the R&S
FSW User Manual.
Remote command:
Single Sweep / Run Single
While the measurement is running, the "Single Sweep" softkey and the RUN SINGLE
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Note:
Sequencer. Furthermore, the RUN SINGLE key controls the Sequencer, not
individual sweeps. RUN SINGLE starts the Sequencer in single mode.
If the Sequencer is off, only the evaluation for the currently displayed measurement
channel is updated.
Remote command:
Continue Single Sweep
After triggering, repeats the number of sweeps set in "Sweep Count", without deleting
the trace of the last measurement.
While the measurement is running, the "Continue Single Sweep" softkey and the RUN
SINGLE key are highlighted. The running measurement can be aborted by selecting
the highlighted softkey or key again.
Remote command:
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
Performing Measurements