Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
127
User Manual 1173.9286.02 ─ 15
Manual operation:
See
CALCulate<n>:EVALuation[:STATe]
<State>
This command turn integration of the measurement range for residual noise calculation
on and off.
Suffix:
<n>
.
Parameters:
<State>
OFF | 0
Calculates the residual noise over the entire measurement
range.
ON | 1
Calculates the residual noise over a customized range.
*RST:
0
Example:
CALC:EVAL ON
Uses a customized offset range for residual noise calculation.
Manual operation:
See
CALCulate<n>:EVALuation:STOP
<OffsetFrequency>
This command defines the end point of the residual noise integration range.
Before you can use the command, you have to turn on the measurement range inte-
gration with
CALCulate<n>:EVALuation[:STATe]
Suffix:
<n>
.
Parameters:
<OffsetFrequency>
The minimum offset is 1
Hz. The maximum offset depends on
the hardware you are using.
*RST:
1 MHz
Example:
CALC:EVAL:STOP 1 MHZ
Defines an end point of 1 MHz for the residual noise range.
Manual operation:
See
CALCulate<n>:EVALuation:USER<range>:STARt
<OffsetFrequency>
This command defines the start point of a custom residual noise calculation range.
Before you can use the command, you have to assign the user range to a trace with
CALCulate<n>:EVALuation:USER<range>:TRACe
Suffix:
<n>
.
Configure Numerical Result Displays