6.2.1 Configuring Residual Noise Measurements
Meas Range
Turns the integration of the entire measurement range for residual noise calculations
on and off.
The range defined here is applied to all traces.
"On"
The application calculates the residual noise over the entire measure-
ment range.
"Off"
The application calculates the residual noise over a customized
range.
The input fields next to the "On/Off" control become available to
define a customized integration range. The application shows two red
lines ("EL1" and "EL2") in the graphical result display to indicate the
custom range.
Remote command:
Turn customized range on and off:
CALCulate<n>:EVALuation[:STATe]
Define start point of custom range:
Define end point of custom range:
User Range
Defines a custom range for residual noise calculations. You have to assign a user
range to a particular trace.
In the default state, user ranges are inactive. "None" is selected in the dropdown
menu. If you assign the user range to a trace by selecting one of the traces from the
dropdown menu, the input fields next to the trace selection become active. In these
fields, you can define a start and stop offset frequency.
Configure Numerical Result Displays