Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
118
User Manual 1173.9286.02 ─ 15
Manual operation:
See
DISPlay[:WINDow]:TRACe<t>:MODE
<Mode>
This command selects the trace mode.
Suffix:
<t>
.
Parameters:
<Mode>
WRITe | VIEW | AVERage | MAXHold | MINHold | BLANk
*RST:
Trace 1/2: WRITe, Trace 3-6: BLANk
Example:
INIT:CONT OFF
SWE:COUN 16
Turns on single sweep mode and defines a count of 16 mea-
surements.
DISP:TRAC2:MODE AVER
Select average trace mode for trace 2.
INIT;*WAI
Performs the measurement (16 sweeps) with synchronization to
the end.
Manual operation:
See
DISPlay[:WINDow]:TRACe<t>:SMOothing:APERture
<Aperture>
This command defines the degree (aperture) of the trace smoothing.
A single aperture applies to all traces which require smoothing.
Suffix:
<t>
.
Parameters:
<Aperture>
Range:
1 to 20
*RST:
0
Default unit: PCT
Example:
DISP:TRAC:SMO:APER 5
Defines an aperture of 5%.
Manual operation:
See
DISPlay[:WINDow]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a particular trace on and off.
Suffix:
<t>
.
Graphical Display of Phase Noise Results