Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
82
User Manual 1173.9286.02 ─ 15
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
Manual operation:
See
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see the "Remote Basics" chapter in the R&S
FSW User
Manual.
Suffix:
<n>
.
irrelevant
Usage:
Event
Manual operation:
See
INITiate<n>:SEQuencer:ABORt
This command stops the currently active sequence of measurements. The Sequencer
itself is not deactivated, so you can start a new sequence immediately using
INITiate<n>:SEQuencer:IMMediate
To deactivate the Sequencer use
Suffix:
<n>
.
irrelevant
Usage:
Event
INITiate<n>:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
command used for a single
measurement.