
Operating the instrument
R&S
®
RTP
59
Getting Started 1803.6478.02 ─ 01
Image
Saves a screenshot of the current display using the settings defined in
"Menu" >"Save/Recall" > "Save" tab > "Screenshot".
Report
Creates a report of the current measurement settings and results using the
settings defined in "Menu" > "Save/Recall" > "Save" tab > "Report".
Clear
Deletes all measurement results including long term measurement and sta-
tistics, all waveforms, and the history.
Autoset and Preset
Performs an autoset, or a preset to a default state. The icons have
the same functionality as the corresponding keys on the front panel. They are
useful when you operate the instrument remotely.
Run / Stop and Single
Starts and stops the continuous acquisition, or starts a defined num-
ber of acquisition cycles. The icons have the same functionality as the corre-
sponding keys on the front panel. They are useful when you control the instru-
ment remotely.
Find Level
Analyses the signal and sets the trigger level to the middle of the signal
peaks.
Force Trigger
Starts an immediate single acquisition. If the acquisition is running in nor-
mal mode and no valid trigger occurs, use "Force Trigger" to confirm that a signal
is available. Then you can use the displayed waveform to determine how to trig-
ger on it.
Demo
Opens the "Demo" dialog, where you can find examples of operating the
R&S
RTP.
Zoom
Creates a zoom. Tapping the icon opens the overlay menu, where you
select one of the four zoom types: standard zoom, hardware zoom, coupled
zoom, fingertip zoom.
Toolbar