Remote Control
R&S
®
NRT2
189
User Manual 1178.5550.02 ─ 04
ON
Switches the relative display mode on and the absolute display
off.
OFF
OFF has the opposite effect.
*RST:
0
Example:
UNIT1:POW:REL:STAT ON
UNIT<Measurement>:POWer:RELative[:VALue]
<unit>
This command allows the forward power (FWD) and the absorbed power (F-R) to be
indicated in relative display mode. For the setting to become effective, the relative dis-
play mode must be switched on, see
UNIT<Measurement>:POWer:RELative:
UNIT<n>:POWer:RELative: STATe ON
command.
Suffix:
<Measurement>
.
1...100
Parameters:
<unit>
PCT | DB
PCT
Deviation from reference value in percent.
DB
Deviation from reference value in dB.
*RST:
PCT
Example:
UNIT2:POW:REL:STAT ON
UNIT2:POW:REL DB
UNIT