Remote Control
R&S
®
NRT2
144
User Manual 1178.5550.02 ─ 04
Connector 1 can be configured as analog output. Analog output may be a DA conver-
sion of a measurement result or a signaling of a limit violation (0 V or 2.5 V).
Connector 2 can be configured as analog output or as trigger input.
Suffix:
<output>
.
1...2
Setting parameters:
<mode>
OFF | RECorder | FRECorder | RRECorder | LIMit | FLIMit |
RLIMit | TOUT | TIN
OFF
The connector is disabled.
FRECorder
The output voltage of the connector follows the forward power of
the measurement.
RRECorder
The output voltage of the connector follows the reflection power
of the measurement.
FLIMit
TTL mode of connector 1. Output is either 0 V or 2.5 V depend-
ing on the measured forward power and the setting of
on page 143 (connector 1 only).
RLIMit
TTL mode of connector 1. Output is either 0 V or 2.5 V depend-
ing on the measured reflection power and the setting of
on page 143 (connector 1 only).
TIN
Connector is configured as trigger input (connector 2 only).
*RST:
OFF
Usage:
Setting only
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio[:VALue]
<value>
Specifies the lower power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power ratio (unit
DB
,
DPCT
or
O
) as the mea-
sured value.
Unit:
DB | DPCT | O
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
1e-18 to 1e18
*RST:
1.0
Default unit: -
Manual operation:
See
OUTPut