Remote Control
R&S
®
NRT2
108
User Manual 1178.5550.02 ─ 04
Manual operation:
See
*SRE
<register>
Service Request Enable
Sets the service request enable register to the specified value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<register>
Range:
0 to 255
*RST:
0
*SRQ?
[<timeout>]
A generic
srq wait
command to be used without
srq event transport
. It is sim-
ply read from the interface.
Query parameters:
<timeout>
Return values:
<srq>
Usage:
Query only
*STB?
STatus Byte query
Returns the contents of the status byte in decimal form.
Usage:
Query only
*TRG
TRiGger
Triggers a measurement. This command is only valid if the power sensor is in the
WAIT_FOR_TRIGGER state and the trigger source is set to
BUS
Usage:
Event
*TST?
Selftest query
Triggers a self test of the instrument and outputs an error code in decimal form.
0
indi-
cates that no errors have occurred.
Common Commands