Preparing for Use
R&S
®
SMC100A
25
Quick Start Guide 1411.4077.62 ─ 07
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting
the power cord, the instrument loses its current settings. Furthermore, program data
may be lost.
Press the ON/STANDBY key first to shut down the application properly.
1. Press the ON/STANDBY key to save the current setup, shut down the operating
system and set the instrument to standby state.
The yellow LED must be on.
2. To switch off the power, press the main power switch to position 0 (Off).
None of the front-panel LEDs should be on.
2.3.7 Power Fuses
The R&S
SMC is fully fused by two fuses IEC60127-T3.15H/250 V.
The fuses are accommodated in the fuse holders next to the power connector. Use
only fuses of the mentioned type.
Shock hazard
Before replacing a fuse, make sure that the instrument is switched off and disconnec-
ted from all power supplies.
Always use fuses supplied by Rohde & Schwarz as spare parts, or fuses of the same
type and rating.
2.4 Connecting External Accessories
The equivalent USB ports on the front and/or rear panel of the R&S
SMC can be used
to connect a variety of accessories.
In addition the instrument provides interfaces for network integration (see
"Setting Up a Network (LAN) Connection"
2.4.1 Connecting USB Devices
Via the USB ports, you can connect external devices directly to the R&S
SMC. If you
need more interfaces, you just use an USB hub.
Connecting External Accessories