![Rohde & Schwarz 1411.4002K02 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/rohde-and-schwarz/1411-4002k02/1411-4002k02_quick-start-manual_1477928030.webp)
Preparing for Use
R&S
®
SMC100A
15
Quick Start Guide 1411.4077.62 ─ 07
LF
Output for internal LF modulation generator signal.
See also data sheet and Operating Manual, section "LF Generator and LF Output".
RF 50 Ohm
Output for RF signal.
NOTICE!
Maximum Input Levels. Do not overload the RF output.
The instrument is equipped with a reverse power protection that prevents the RF out-
put against back feed, see section "Reverse Power Protection" in the operating man-
ual. However, the maximum permissible reverse power is specified in the data sheet.
The RF signal is output via an N female connector.
NOTICE!
Risk of RF connector and cable damage. Excessive tightening of the connec-
tors can damage the cables and connectors. Too weak tightening leads to inaccurate
results.
Always use an appropriate torque wrench suitable for this type of connector and apply
the torque specified in the application note
The application notes are available on the Internet and provide additional information
on care and handling of RF connectors.
Rohde
&
Schwarz offers appropriate torque wrenches for various connectors. For
ordering information see the R&S
SMC data sheet or product brochure.
2.2 Rear Panel Tour
This section gives an overview of connectors on the rear panel of the instrument. Each
connector is briefly described and a reference is given to the chapters containing
detailed information. For technical data of the connectors refer to the data sheet.
For more detailed information, refer to the corresponding description in the Operating
Manual.
Rear Panel Tour