GUI Reference
R&S
®
ZNA
526
User Manual 1178.6462.02 ─ 12
Remote command:
[SENSe<Ch>:]PATH<Pt>:DIRectaccess
SOURce<Ch>:PATH<Pt>:DIRectaccess
[SENSe<Ch>:]PATH<Pt>:MEASurement:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:REFerence:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
Rear Out
If checked, the
connectors on the rear panel are used as (additional)
output.
Remote command:
[SENSe<Ch>:]PATH<Pt>:IFSWitch
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
Auto Determine IF Out
This checkbox is only enabled, if
is used.
If unchecked (default), you can define the
manually. However, this
frequency is also used internally and is possibly not optimal for the Direct IF Access
Input / Output ports. If you activate "Auto Determine IF Out", the analyzer firmware
selects suitable rear IF frequencies.
Remote command:
[SENSe<Ch>:]PATH<Pt>:IFOutauto
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFOutauto?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFOutauto?
Rear Input Path
This selection is only enabled if "Rear IN" is used as
different paths are available for the IF signal.
●
"Direct" – frequency range 5kHz to 1 GHz
●
"LP_60M" – low pass, frequency range 5 kHz to 60 MHz
●
"BP_279M" – band pass, frequency range 260 MHz to 290 MHz
The limitation of the Frequency to the smaller frequency ranges increases the system
performance. Standard case could be a fixed frequency input, as it is for the mm-con-
verter IF frequency.
Remote command:
[SENSe<Ch>:]PATH<Pt>:IFINpath
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFINpath?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFINpath?
Rear IF Frequency
Sets the frequency at the related
ports as a linear transformation of
the channel base frequency f
b
.
This setting is available in one of the following cases:
●
"Rear IN" is used as
●
is used and
Channel Config Softtool