Command Reference
R&S
®
ZNA
1168
User Manual 1178.6462.02 ─ 12
Make sure the converted frequency or frequency range is within the Direct IF Access
Input / Output ports' frequency range. If the resulting frequencies are not (completely)
within the allowed range, the analyzer returns an error message. Nevertheless, the fre-
quency formula is applied.
Note:
This setting only takes effect in one of the following cases:
●
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
IN
.
●
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
OUT
and
[SENSe<Ch>:]PATH<Pt>:IFOutauto
OFF
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Setting parameters:
<Numerator>
The specified numerator is rounded to the next higher integer;
zero is not allowed.
<Denominator>
The specified denominator must be positive. It is rounded to the
next higher integer.
<Offset>
The specified offset value is rounded to (positive or negative)
multiples of 1 Hz.
<Sweep type>
FB | CW
FB
The IF frequency at IF Reference <Pt> and IF Meas <Pt>
sweeps with the channel base frequency.
CW
Fixed IF frequency at IF Reference <Pt> and IF Meas <Pt>.
*RST:
SWEep
Usage:
Setting only
Options:
R&S
ZNA-B26
Manual operation:
See
[SENSe<Ch>:]PATH<Pt>:IFSWitch
<Selection of IF Switch on Ref and Meas path>
Defines if and how the Direct IF Access Input / Output ports IF Reference <Pt> and IF
Meas <Pt> are used in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Setting parameters:
<Selection of IF
Switch on Ref and
Meas path>
OFF | IN | OUT
OFF
The ports are unused.
SCPI Command Reference