Command Reference
R&S
®
ZNA
1167
User Manual 1178.6462.02 ─ 12
BP_279m
Band pass, frequency range 260 MHz to 290 MHz
Usage:
Setting only
Options:
R&S
ZNA-B26
Manual operation:
See
[SENSe<Ch>:]PATH<Pt>:IFOutauto
<Boolean>
Defines how the outgoing frequencies at IF Reference <Pt> and IF Meas <Pt> are
determined in channel <Ch>.
This setting only takes effect if
OUT
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Setting parameters:
<Boolean>
OFF
The frequencies are set manually.
Use
[SENSe<Ch>:]PATH<Pt>:IFRequency
ble frequency conversion formula.
ON
The frequencies are set automatically (by the analyzer firm-
ware).
*RST:
OFF
Usage:
Setting only
Options:
R&S
ZNA-B26
Manual operation:
See
[SENSe<Ch>:]PATH<Pt>:IFRequency
<Numerator>, <Denominator>, <Offset>,
<Sweep type>
Defines the freqency conversion formula for Direct IF Access Input / Output ports IF
Reference <Pt> and IF Meas <Pt> in channel <Ch>.
For <Sweep type>
FB
, the IF frequency f
IF
is calculated according to
f
IF
= <Numerator>/<Denominator> * f
b
+ <Offset>,
where f
b
represents channel base frequency.
For <Sweep Type> CW, f
IF
is calculated according to
f
IF
= <Offset>,
i.e. f
IF
is constant.
SCPI Command Reference