Command Reference
R&S
®
ZNA
1074
User Manual 1178.6462.02 ─ 12
<L0>, <L1>,
<L2>, <L3>
mandatory
*
mandatory
mandatory
*
mandatory
**
mandatory
**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional
***
optional
***
optional
***
mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1>
optional: <Port 1>,<Port 2>
*
values will be ignored during calibration
**
if OPEN is selected the residual inductance (L0,...,L3) will be ignored during calibration; if SHORt is selected the fringing capaci-
tance (C0,...,C3) will be ignored during calibration
***
must be provided if a port restriction <Port 1> shall be applied; values will be ignored during calibration
The parameters in the
[SENSe<Ch>:]CORRection:CKIT:<StandardType>
,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels
, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType>
com-
mands have the following meaning:
Table 8-10: Parameter list
Parameter
Meaning
Comment/Unit
'<ConnType>'
Name of the connector type.
Use
SENSe1:CORRection:CONNection:CATalog?
to query connec-
tor names.
String parameter
'<CalKitName>'
Name of the calibration kit. Use
[SENSe:]CORRection:CKIT:CATalog?
to query cal kit names.
String parameter
'<CalKitLabel>'
Label (e.g. the serial number) of the calibration kit; for
...WLABel
com-
mand only
String parameter
'<StandardLa-
bel>'
Label (e.g. the serial number) of the standard
String parameter
<MinFreq>,
<MaxFreq>
Min./max. frequency for which the circuit model is valid
Default unit is Hz
<DelayParam>
Depending on the cal kit's model type (selected using
) this parameter is
interpreted as
●
delay [s] for Agilent modelling
●
el. length [m] for ZVR compatible modelling
Cal kits that are not created/modified using
[SENSe:]CORRection:CKIT:DMODe
use the ZVR compatible
modelling.
<Loss>
Loss (offset parameter) of the standard
To be specified without unit (implicit
unit is dB)
<Z0>
Reference impedance (no unit)
To be specified without unit (implicit
unit is Ω)
<C0>,...,<C3>
Polynomial coefficients for the fringing capacitance of the standard (load
parameter)
To be specified without unit: implicit
unit of <C
i
> is fF / (GHz)
i
<L0>,...,<L3>
Polynomial coefficients for the residual inductance of the standard (load
parameter)
To be specified without unit: implicit
unit of <L
i
> is pH / (GHz)
i
SCPI Command Reference