Command Reference
R&S
®
ZNA
1068
User Manual 1178.6462.02 ─ 12
[SENSe:]CORRection:CKIT:CATalog?
[<ConnectorType>]
Returns a list of all cal kits for a given connector type or for all connector types.
Query parameters:
<ConnectorType>
Name of the connector type (optional). Use
]CORRection:CONNection:CATalog?
names.
If omitted, the command returns the list of all cal kits.
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
Manual operation:
See
[SENSe:]CORRection:CKIT:DELete
<CalKitName>
Deletes an imported or user-defined cal kit.
Note:
It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
Example:
See
Usage:
Setting only
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:DMODe
<ConnectorType>, <CalKitName>, <KitLabel>,
<arg3>
Sets/gets the delay mode for the related cal kit (identified by connector type, name and
label), i.e allows to toggle between ZVR compatible and Agilent modelling (see
on page 481). Subsequent standard definitions interpret the specified
<DelayParam> accordingly.
In "set direction", if a cal kit with the given connector type, name and label is not availa-
ble on the analyzer, it is created automatically.
Cal kits that are not created/modified using
[SENSe:]CORRection:CKIT:DMODe
use
the ZVR compatible modelling.
Parameters:
<ConnectorType>
Name of the connector type. Use
]CORRection:CONNection:CATalog?
names.
<CalKitName>
String parameter containing the name of the calibration kit. See
"Cal kit naming conventions " under
SCPI Command Reference