With the sweep type "Sweep", only swept frequency sweeps can be selected;
for the sweep type "FFT", only FFT sweeps can be selected. In the FSP, the
FFT mode was under the "Filter type" softkey. In the FSV, this setting is now
under "Sweep type".
●
The HELP key provides context sensitive online help including remote com-
mands.
●
After entering an application once (like analog demodulation), a second tab is
opened in the upper part of the display. This allows easy switching between
applications simply by touching the tabs.
●
16 markers are supported and a marker table is available. If tables get too
large you can scroll using the touchscreen or maximize the table using the
CHANGE FOCUS and MAXIMIZE/SPLIT keys.
●
A wizard is available in the TRACE menu to easily set up all 6 traces at a
glance.
●
Dialog boxes and input fields are on the upper right side of the display to be
near the softkeys. They can be moved to any place on the screen and re-
appear there when re-opened.
●
An FSP-compatible mode can be selected. In this case, the FSV behaves as
an FSP (e.g. same number of sweep points, bandwidths like FSP, etc.) and
also identifies itself at the *IDN command as an FSP. Thus, FSP remote pro-
grams can be re-used.
●
Under "Setup > Display Setup" various themes for the color definition can be
found, including a "GrayStone" theme that makes the FSV look like it has FSP
softkeys.
●
USB connectors are available on the front.
●
The instrument account password has changed from the FSP to FSV. The
account password is no longer
instrument
but rather
894129
. Thus, you can
easily retrieve local instrument control after Remote Desktop usage via the
instrument's number pad.
●
There is a link on the Windows Desktop and in the "Start" menu to access the
R&S user data. The link points to the standard directory of hardcopies and
save/recall files, so if the files are to be copied on a memory stick they can be
found easily.
●
The FSP cold boot feature with the decimal point no longer exists. Instead, the
shutdown and calibration files can be deleted via the link "Delete Shutdown
Files" in the "Start > All Programs" menu, if necessary.
●
The optional external mixer functionality has been improved to provide more
flexibility and conveniency, in particular:
New Features for R&S
FSP Users