Controlling the R&S
FSVA/FSV Remotely
R&S
®
FSVA/FSV
144
Quick Start Guide 1321.3066.02 ─ 06
Configuring the Power Save Function for the Display
During remote control operation, it is often unnecessary to display the measure-
ment results on screen. Although the command
SYSTem:DISPlay:UPDate
OFF
switches off the display of the measurement results, thus significantly
improving speed in remote control operation, the display itself and the back-
ground lighting in particular remain switched on.
If you also want to switch off the display itself, you must use the power save func-
tion by setting the response time in minutes prior to activation.
The display is reactivated immediately when you press a key on the instru-
ment's front panel.
Public SUB PowerSave()
CALL InstrWrite(analyzer, "DISPlay:PSAVe:HOLDoff 1")
'Set response time to 1 minute
CALL InstrWrite(analyzer, "DISPlay:PSAVe ON")
'Switch on Power Save function
8.3.1.3
Sending Simple Instrument Setting Commands
This example shows how the center frequency, span and reference level of the
instrument are set.
REM -------- Instrument setting commands -----------------------------------
PUBLIC SUB SimpleSettings()
CALL InstrWrite(analyzer, "FREQUENCY:CENTER 128MHz")
'Center frequency 128 MHz
CALL InstrWrite(analyzer, "FREQUENCY:SPAN 10MHZ")
'Set span to 10 MHz
CALL InstrWrite(analyzer, "DISPLAY:TRACE:Y:RLEVEL -10dBm")
'Set reference level to -10dBm
END SUB
REM ************************************************************************
8.3.1.4
Switching to Manual Operation
REM -------- Switching instrument to manual operation ----------------------
CALL viGpibControlREN(analyzer, VI_GPIB_REN_ADDRESS_GTL)
'Set instrument to Local state
REM ************************************************************************
Brief Introduction to Remote Control