Chapter 2: Control Testing
ACCU-CHEK Customer Care Service Center 1-800-858-8072
3.
Briefly roll the bottle of control solution, open it, squeeze one drop onto
a paper towel, and wipe the tip of the bottle clean.
4.
Touch and hold one drop of control solution to the edge of the yellow
window. When the test is done, the meter displays the test result.
5.
Check that the Level 1 result is within the Level 1 range on the
test strip vial.
Run Test
08-07-03 10:00AM
59
mg/dL
MENU
DIARY
EVENT
If the result is not within
the acceptable range, see
Understanding Control
Test Results on page 21.
Example
Range (mg/dL)
Level 1 (LO)
42–72
Level 2 (HI)
290–328
Example
Example
6.
Press
EVENT
, then take the test strip out and discard it.
Make sure you press
EVENT before
you take the strip out and
while the test result is still displayed on the screen.