General Technical Data
Barometric pressure
During operation: 55 to 106 kPa (550 to 1060 mbar)
During storage in its shipping case: 70 to 106 kPa (700 to 1060 mbar)
During transport 50 to 106 kPa (500 to 1060 mbar)
Power supply
One 1.5 V AAA Alkaline (LR03) or Lithium (FR03) battery. Batteries should have a
minimum capacity of 1200 mAh. Do not use carbon zinc or rechargeable (NiCd/NiMH)
batteries.
Lifetime of battery
If used in a typical usage pattern (50 U/day using U100 insulin; operating temperature
23 °C ±2 °C [73 °F ±4 °F]), the battery life is approximately:
Alkaline (1200 mAh):
5 days with
Bluetooth
off
Lithium (1200 mAh):
14 days with
Bluetooth
off
Data storage time
The time and date are safely stored in the memory for about 10 hours after the battery
has been removed. Other insulin pump settings (the hourly basal rates, remaining
cartridge content, bolus increments, and active user profile) and the event memory (bolus
history, history of daily insulin totals, Temporary Basal Rate history, alarm history) are
saved, regardless of battery power and the time your insulin pump has been without a
battery.
Содержание Accu-Check Insight
Страница 3: ...Pump details and accessories Adapter tubing Battery Battery cover Cartridge Pump...
Страница 4: ......
Страница 8: ......
Страница 16: ...6 1 1 Navigating through the menus q Quick Bolus Keys Display Screen x Up Key o OK Key z Down Key l Zoom Key...
Страница 62: ...52...
Страница 76: ...66...
Страница 96: ...86...
Страница 140: ...130...
Страница 146: ...136...
Страница 159: ...149 Appendices...
Страница 160: ...150...
Страница 181: ...C 171...
Страница 185: ...D 175 Icon Meaning Radio button selected Radio button not selected Icon on Icon off...
Страница 188: ...178 Symbol Meaning Warning against device use in electro magnetic or high electrical field or environments...
Страница 205: ...195...
Страница 211: ...201...