8.4.4 Adjusting the Cartridge warning level
When the insulin amount in your pump’s cartridge is low your
pump triggers a
Warning W31: Cartridge low
to remind you that
you need to change the cartridge soon.
You can specify the number of insulin units remaining in your
cartridge that will trigger the Warning.
1
From the
Settings
menu, press
z
to move
to
Therapy settings
.
Press
o
to select.
2
The
Therapy settings
menu appears.
Press
z
to move to
Cartridge warning
level
.
Press
o
to select.
3
The
Cartridge warning level
screen
appears.
Press
x
or
z
to set the warning amount
as needed.
Press
o
to confirm.