10
145
Press
o
to mute
Press
o
to confirm
Maintenance message
M24 Occlusion
:
The insulin cannot flow freely. Change the complete infusion
set. If the Maintenance message occurs again, change the
cartridge as well.
M25 Loantime over
(loan pumps only):
The loan time of your pump has expired. Contact your local
pump support (see inside back cover) to discuss appropriate
action. After a Maintenance message M25 the pump will be in
STOP mode.
M26 Cartridge change not completed
:
The cartridge change process was interrupted. Repeat the
cartridge change.
M27 Data download failed
:
The data transfer to a computer has been interrupted. Restart
the data download.
M28 Pause mode timeout over
:
Your pump has been in PAUSE mode for more than 15 minutes
and has switched to STOP. TBR and boluses have been
cancelled. Restart insulin delivery and reprogram TBR and
boluses as appropriate.
Содержание Accu-Check Insight
Страница 3: ...Pump details and accessories Adapter tubing Battery Battery cover Cartridge Pump...
Страница 4: ......
Страница 8: ......
Страница 16: ...6 1 1 Navigating through the menus q Quick Bolus Keys Display Screen x Up Key o OK Key z Down Key l Zoom Key...
Страница 62: ...52...
Страница 76: ...66...
Страница 96: ...86...
Страница 140: ...130...
Страница 146: ...136...
Страница 159: ...149 Appendices...
Страница 160: ...150...
Страница 181: ...C 171...
Страница 185: ...D 175 Icon Meaning Radio button selected Radio button not selected Icon on Icon off...
Страница 188: ...178 Symbol Meaning Warning against device use in electro magnetic or high electrical field or environments...
Страница 205: ...195...
Страница 211: ...201...