A
159
Table 203: Immunity electromagnetic fields (life-supporting devices)
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Insight insulin pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the Accu-Chek Insight insulin pump should assure that it is used in such an environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Accu-Chek Insight insulin pump,
including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the
transmitter.
Immunity test standard
IEC 60601 test level
Compliance level
Recommended separation distance c)
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
V1 = 10 Vrms 150 kHz
to 80 MHz
d = 1.2√
−
P
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 800 MHz
E1 = 10 V/m 80 MHz
to 800 MHz
d = 1.2√
−
P
80 MHz to 800 MHz
Radiated RF
IEC 61000-4-3
10 V/m
800 MHz to 2.5 GHz
d)
E2a = 2 V/m for
2.3 GHz to
2.5 GHz
e)
E2b = 10 V/m for
800 MHz to
2.3 GHz and
2.5 GHz to
2.7 GHz
da = 11.7√
−
P
2.3 GHz to 2.5 GHz
db = 2.3√
−
P
800 MHz to 2.3 GHz and
2.5 GHz to 2.7 GHz
Содержание Accu-Check Insight
Страница 3: ...Pump details and accessories Adapter tubing Battery Battery cover Cartridge Pump...
Страница 4: ......
Страница 8: ......
Страница 16: ...6 1 1 Navigating through the menus q Quick Bolus Keys Display Screen x Up Key o OK Key z Down Key l Zoom Key...
Страница 62: ...52...
Страница 76: ...66...
Страница 96: ...86...
Страница 140: ...130...
Страница 146: ...136...
Страница 159: ...149 Appendices...
Страница 160: ...150...
Страница 181: ...C 171...
Страница 185: ...D 175 Icon Meaning Radio button selected Radio button not selected Icon on Icon off...
Страница 188: ...178 Symbol Meaning Warning against device use in electro magnetic or high electrical field or environments...
Страница 205: ...195...
Страница 211: ...201...