9
Measuring L/C/R
Measuring D/Q/
θ
/DCR
Discharge the DUT (Device Under Test) before connecting the test probes.
The figures on the following pages show how to make basic measurements.
The auto test mode is the default mode when the meter is turned on. When the
meter in auto test mode, it will automatically detect the DUT and show the suitable
result on the dis play.
The meter can measure D (dissipation factor), Q (quality factor), θ (phase) and
DCR (DC resistance) on the DUT.
• Press the D/Q/θ/DCR button to enter D/Q/θ/DCR test mode.
• In D/Q/θ/DCR test mode, press the D/Q/θ/DCR button to select measuring function.
• In auto test mode, press the AUTO L/C/R button to enter manual test mode.
• In manual test mode, press the AUTO L/C/R button to select measuring function.
• To return the auto test mode, press and hold the AUTO L/C/R button > 2 sec.