16
General Specifications
Inductance
D & Q
θ
Definition: Q = 1/D = tanθ
Range: 2.000 to 2000
Minimum Resolution: 0.001
Accuracy: ± (0.5% + 5) x (1+D), when D < 1 or Q > 1
Definition: θ =
tan
-1
Q
Range: -90.0° to 90.0°
Minimum Resolution: 0.1°
Accuracy: ± (0.5% + 5)
Range
100Hz
1kHz
10kHz
120Hz
20.000uH
N/A
N/A
N/A
200.00uH
N/A
N/A
0.5% + 30
[1]
2000.0uH
N/A
0.5% + 30
[1]
0.5% + 5
20.000mH
0.5% + 30
[1]
0.5% + 5
0.5% + 5
200.00mH
0.5% + 5
0.5% + 5
0.5% + 5
2000.0mH
0.5% + 5
0.5% + 5
1.0% + 5
[2]
[1] Accuracy is specified after subtract of the offset inductance.
[2] < 50dgt rolling.
[3] If D > 0.1, the accuracy should be multiplied by 1 + D