![RND 365-00002 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/rnd/365-00002/365-00002_user-manual_1467543017.webp)
17
Capacitance
Range
100Hz
1kHz
10kHz
120Hz
200.00pF
N/A
N/A
2.0% + 1pF
[1]
2000.0pF
0.5% + 8
[1]
0.5% + 8
[1]
0.5% + 8
[1]
20.000nF
0.5% + 5
0.5% + 5
0.5% + 5
200.00nF
0.5% + 5
0.5% + 5
0.5% + 5
2000.0nF
0.5% + 5
0.5% + 5
1.0% + 5
20.000uF
0.5% + 5
1.0% + 5
N/A
200.00uF
1.0% + 5
N/A
N/A
[1] Accuracy is specified after subtract of the stray capacitances for test leads.
[2] < 50dgt rolling.
[3] If D > 0.1, the accuracy should be multiplied by 1 + D
2
Resistance
Range
100Hz
1kHz
10kHz
120Hz
20.000Ω
N/A
0.5% + 50
[1]
0.5% + 50
[1]
200.00Ω
0.5% + 8
[1]
0.5% + 8
[1]
0.5% + 8
[1]
2.0000kΩ
0.5% + 5
0.5% + 5
0.5% + 5
20.000kΩ
0.5% + 5
0.5% + 5
0.5% + 5
200.00kΩ
0.5% + 5
0.5% + 5
1.0% + 5
[2]
2.0000MΩ
0.5% + 5
1.0% + 5
[2]
N/A
20.000MΩ
1.0% + 5
[2]
N/A
N/A
[1] Accuracy is specified after subtract of the offset resistance.
[2] < 50dgt rolling.
[3] If Q > 0.1, the accuracy should be multiplied by 1 + Q
2
Содержание 365-00002
Страница 6: ...6 Measuring Principle Series Measuring ...
Страница 23: ...23 Messprinzip Serienmessung ...
Страница 39: ...39 Montering 1 USB kabel 2 Nätadapter 3 L typ testprober och specialskruvar x 4 4 CD med Programvara 1 2 3 4 ...
Страница 40: ...40 Mätprincip Serie Mätning ...
Страница 53: ...Distrelec Group AG Grabenstrasse 6 CH 8606 Nänikon ...