40
As a policy of continual improvement, RMF reserves the right to alter the specification without prior notice.
201.028 REV 1
Date of Issue: 18 June 2018
AS4059E Table 2 uses letters instead of numbers to indicate the particle size range, so the settings are labelled
appropriately. The standard specifies ways to represent a cleanliness level using only a subset of the available
particle sizes, for example B-F. The user can achieve this by only entering settings for the sizes desired, leaving the
others empty. So a limit of AS4059 7B-F could be represented simply by entering a value of 7 for B, C, D, E and F.
AS4059E Table 1 / ISO11218 Alarm Levels
Figure 6.27
These two standards are similar except for terminology and reporting format. The actual numeric sizes and class
thresholds are the same.
6.3.3.9.2 Alarm Mode
Figure 6.28
The Alarm Mode sets the precise function of the two switched alarm outputs of the CMS.
This allows the CMS to be used in a variety of situations. Note that the conditions under which the outputs are
turned on are also displayed above the Alarm Mode selector, for each setting.
NOTE: These outputs are distinct from the front panel LED, and that the set alarm mode does not affect the LED.
The set alarm mode determines the function of the two switched outputs only. This setting and this entire section
can be ignored if these outputs are unused, i.e. the user has not connected them to anything.
New modes are occasionally added after a customer request, this means that modes may not all be implemented
unless using the latest firmware revision.