Connection of a water meter for analyses at volume intervals
In addition to the time-dependent triggering of an analysis, there is also the possibility of quantity-
dependent triggering. For this, either a contact water meter with a potential-free reed switch or an
electronic turbine with a Hall switch is used.
(When using this function, the Sycon must be in automatic mode).
Programming: Menu > Settings > Inputs > Input 1 > Water meter
1.
Enter the characteristic data of the water meter used in the unit
litre/pulse
.
2.
Enter the type of water meter Hall or Reed. In Reed mode, the input is debounced to prevent
miscounting.
Programming: Menu > Settings > Analysis > Auto.Interval Quantity
3.
Activate analyses in the quantity interval.
4.
Select the unit of measure to be shown in the display.
5.
Enter the quantity after which an analysis is to take place.
After switching on the unit, the first analysis is started after 3 minutes. The
water quantity count is already active during this time.
The operation of contact water meters (Reed) in the input mode for
electronic water meters (Hall) can lead to incorrect interval quantities due to
contact bounce.
Programming: Menu > Settings > Inputs > Input 1 > Water meter
Programming: Menu > Settings > Analysis > Auto.Interval Quantity
Figure 18Three possible connection variants around the analyser (2)
Connection of external components