background image

 

RP-2009 Pump for the GX-2009

  7

Optional Hose & Probe

When desired, the rubber sample nipple may be removed and a sample 
hose and probe with GX-7 fittings (optional accessories) may be connected 
to the inlet fitting. Sample hose lengths are available from 10 feet to 30 feet 
(see Parts List at the end of the manual).

CAUTION:

Sample hose lengths of more than 30 feet are not recommended for 
the RP-2009 because of flow rate reduction.

Normal Operation

1. Start up the GX-2009 and enter normal operation as described in the 

GX-2009 Operator’s Manual.

2. Push down on the RP-2009’s battery cap and rotate it 1/8 turn 

counterclockwise so that the indicator tab on the battery cap is next to 
“RELEASE”. 

Figure 2: Releasing the Battery Cap

WARNING:Do not install the GX-2009 onto the RP-2009 without rotating 

the battery cap to “RELEASE”. Installing the GX-2009 onto 
the RP-2009 with the battery cap in the “LOCK” position may 
result in damage to the RP-2009.

3. Be sure the front labels on the GX-2009 are facing the same way as the 

front labels on the RP-2009. The retaining tabs on the RP-2009 are keyed 
to allow installation of the GX-2009 only if the labels on both 
instruments are facing front. Do not force the GX-2009 into the RP-2009 
if both labels are not facing front.

"RELEASE"

Indicator Tab

S

LO

C

E

P

U

S

H

E

P

O

N

Содержание RP-2009

Страница 1: ...www rkiinstruments com RP 2009 Pump for the GX 2009 Operator s Manual Part Number 71 0267RK Revision P4 Released 10 26 18...

Страница 2: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Страница 3: ...3 Replacing the Hydrophobic Disk Filter and Wire Mesh Disk 14 Parts List 15 WARNING Understand manual before operating and for proper battery type Substitution of components may impair intrinsic safet...

Страница 4: ...able 1 lists specifications for the RP 2009 pump Table 1 Specifications Power Supply 1 ea AA size alkaline Operating Time 8 hours minimum at 25 C Operating Temperature Humidity 20 C to 50 C 0 85 Relat...

Страница 5: ...e 1 RP 2009 Pump Component Location Top Power Button Low Flow LED Power LED Instrument Release Indicator Sensor Adapter Inlet Fitting GX 7 Type Back Front Buzzer Opening Instrument Lock Indicator Pump...

Страница 6: ...pump unit on and off The buzzer opening allows the buzzer s sound to emanate from the buzzer The power LED is on when the pump unit is on The low flow LED flashes when the pump is experiencing a low f...

Страница 7: ...wn on the RP 2009 s battery cap and rotate it 1 8 turn counterclockwise so that the indicator tab on the battery cap is next to RELEASE Figure 2 Releasing the Battery Cap WARNING Do not install the GX...

Страница 8: ...GX 2009 Be sure that the RP 2009 is fully engaged on the instrument When fully engaged the GX 2009 should not move around at all CAUTION If the RP 2009 is not fully engaged the unit will not respond...

Страница 9: ...K Figure 4 Locking the Battery Cap 6 Attach the rubber sample nipple or an optional hose and probe to the inlet fitting Figure 5 Connecting the Optional Hose Probe 7 Attach the RP 2009 GX 2009 assembl...

Страница 10: ...GX 2009 If an optional hose and probe are used instead of the rubber sample nipple hold the probe in the area to be sampled To sample from a well or other similar sample area the hose and probe may be...

Страница 11: ...at the indicator tab on the battery cap is next to RELEASE Figure 6 Releasing the Battery Cap WARNING Do not remove the GX 2009 from the RP 2009 without rotating the battery cap to RELEASE Removing th...

Страница 12: ...16 Press the bottom of the release tab toward the pump unit and pull up on the GX 2009 Figure 7 GX 2009 Removal LOW FLOW POWER RESET Indicator Tab LOCK RELEASE OPEN AIR POWER MODE H2S CO LEL O2 RP 200...

Страница 13: ...lease tab Replacing the Cotton Dust Filter 1 Locate the clear plastic filter holder at one end of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise...

Страница 14: ...d of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise 3 Pull the filter holder away from the pump unit Do not lose the cotton dust filter 4 The hy...

Страница 15: ...it snaps into place Parts List Table 2 lists replacement parts and accessories for the RP 2009 pump Table 2 Parts List Part Number Description 07 2005RK Gasket filter screen retaining 07 6027RK O rin...

Отзывы: