RP-2009 Pump for the GX-2009
7
Optional Hose & Probe
When desired, the rubber sample nipple may be removed and a sample
hose and probe with GX-7 fittings (optional accessories) may be connected
to the inlet fitting. Sample hose lengths are available from 10 feet to 30 feet
(see Parts List at the end of the manual).
CAUTION:
Sample hose lengths of more than 30 feet are not recommended for
the RP-2009 because of flow rate reduction.
Normal Operation
1. Start up the GX-2009 and enter normal operation as described in the
GX-2009 Operator’s Manual.
2. Push down on the RP-2009’s battery cap and rotate it 1/8 turn
counterclockwise so that the indicator tab on the battery cap is next to
“RELEASE”.
Figure 2: Releasing the Battery Cap
WARNING:Do not install the GX-2009 onto the RP-2009 without rotating
the battery cap to “RELEASE”. Installing the GX-2009 onto
the RP-2009 with the battery cap in the “LOCK” position may
result in damage to the RP-2009.
3. Be sure the front labels on the GX-2009 are facing the same way as the
front labels on the RP-2009. The retaining tabs on the RP-2009 are keyed
to allow installation of the GX-2009 only if the labels on both
instruments are facing front. Do not force the GX-2009 into the RP-2009
if both labels are not facing front.
"RELEASE"
Indicator Tab
S
LO
C
E
P
U
S
H
E
P
O
N