RP-2009 Pump for the GX-2009
3. Pull the filter holder away from the pump unit.
4. Remove the old cotton dust filter from the filter holder and replace it
with a new cotton dust filter.
5. Reinstall the filter holder. Align the two tabs on the bottom of the filter
holder with the two slots in the pump unit. Push the filter holder into
the pump unit and turn it 1/4 turn clockwise until it snaps into place.
Replacing the Hydrophobic Disk Filter
and Wire Mesh Disk
1. Locate the clear plastic filter holder at one end of the pump unit.
2. Grasp the ridged sides of the filter holder and turn it 1/4 of a turn
counterclockwise.
3. Pull the filter holder away from the pump unit. Do not lose the cotton
dust filter.
4. The hydrophobic disk filters and wire mesh disk are located in the
pump unit and are retained by a rubber gasket. Pull out the rubber
gasket with needle nose pliers.
Figure 8: Changing the Hydrophobic & Wire Mesh Filters
5. Remove the old hydrophobic filters and/or wire mesh disk from the
gasket. A hydrophobic filter is located on either side of the wire mesh
disk.
POWER
RESET
Rubber Gasket
Installation
Stub
LOW FLOW
POWER
LOCK
Cotton Dust Filter
Hydrophobic Disk Filter
OPEN
RP-2009
RELEASE
Wire Mesh Disk
Filter Holder