64 • Using Setup Mode
GX-2003 Operator’s Manual
1. From the main menu, place the cursor next to the DATALOG
OVERWRITE
menu item. Press and release the POWER ENTER
button.
2. Use the
▲
AIR or (SHIFT)
▼
button to display the desired setting.
3. Press and release the POWER/ENTER button to save the setting
and return to the main menu.
Updating the Date and Time Settings
1. From the main menu, place the cursor next to the DATE/TIME
menu item. Press and release the POWER ENTER button. The date
and time will be displayed with the year flashing.
2. Use the
▲
AIR or (SHIFT)
▼
button to display the desired year.
3. Press and release the POWER ENTER button to save the setting.
The month setting flashes.
4. Repeat steps 2 and 3 to enter the month, day, hours, and minutes
settings. The main menu displays after you enter the minutes
setting.
Updating the Calibration Interval
This setting defines the amount of time between calibrations. The
minimum setting is 1 day and the maximum setting is 255 days. The
factory setting is 90 days.
1. From the main menu, place the cursor next to the CAL.
INTERVAL
menu item. Press and release the POWER ENTER
button.
DATA LOGGER
OVERWRITE
ON
21:09
SET DATE/TIME
Apr/01/’03
21:09
21:09
SET CALIBRATION
INTERVAL
90 DAY
21:09