36 • Display Mode
GX-2003 Operator’s Manual
2. Press and release the
▲
AIR button to confirm that you want to
clear the data logger.
The GX-2003 displays
CLEARING THE DATA LOGGER PLEASE
WAIT . . . .
for several seconds, then displays
CLEARED OK
. The
data logger is cleared and the remaining log time value is reset.
Remaining Log Time Screen
The remaining log time screen displays the time remaining until the
data logger memory is full. The remaining log time depends on how
often the GX-2003 stores data to the data log, how many channels are
active, and how often the GX-2003 is turned on and off.
The table below illustrates how much data logging time is available
for the various interval times. It assumes that the unit is setup with all
5 sensors, is only turned on once, and there are no alarm occurrences.
See “Updating the Interval Time Setting” on page 63 for instructions
on setting the data logging interval time.
Press and release the DISPLAY(ADJ) button once more to return to
Measuring Mode.
Table 5: Data Logging Capacity, 5 Sensor GX-2003
Interval
Time
Data Logging Time
10 seconds
10 hours
20 seconds
20 hours
30 seconds
30 hours
1 minute
60 hours
3 minutes
180 hours
5 minutes
300 hours
10 minutes
600 hours
LOG TIME
9.9
HOURS
REMAINING
21:09