
GX-2003 Operator’s Manual
Start Up, Normal Mode • 21
4. If LUNCH BREAK is turned on (see page 62), the Resume Datalog
Screen displays.
• Press and release the
▲
AIR button to continue accumulating
time-weighted average (TWA), PEAK readings, and time in
operation from the last time the GX-2003 was used. The short-
term exposure limit [STEL] reading is reset each time the GX-
2003 is turned on.
• Press and release the DISPLAY(ADJ) button to reset the
accumulation of these measurements.
If you do not press the
▲
AIR or DISPLAY(ADJ) button within 5
seconds, the GX-2003 automatically resumes accumulating the
TWA, PEAK readings, and time in operation.
5. If USER/STATION ID is turned on (see page 62), the ID Screen
displays for a few seconds.
6. After the ID Screen, if CAL. TIME REMAINING is turned on, the
screen that appears next depends on how CAL.EXPIRED ACTION
is set in the Setup Mode Menu(page 62).
• If the unit is due for calibration and CAL. EXPIRED ACTION
is set to CONFIRM TO USE, then the following screen
displays.
The alarm LED’s and buzzer will pulse several times and the
vibrator will activate for a second. After this, press the RESET
SILENCE button until you hear a beep to continue.
RESUME DATALOG ?
4
YES: AIR
NO: DISPLAY
21:09
USER ID
*********
STATION ID
*********
CAL. DATE PAST
CAL. IS REQUIRED