47
•
Sensor Settings and Calibration
AirLink 6940XP Operator’s Manual
3. Press and release
ADD
or
SUB
until the Background (Background High for O
2
versions) is
set to the desired level.
For non-O2 sensor assemblies, skip to Step 5.
4. For O
2
sensor assemblies, press
ADD
to navigate from the Background High menu item to the
Background Low menu item. Press
MENU
to select the Background Low menu item.
a. Press
ADD
or
SUB
until the Background Low is set to the desired level.
b. Press
MENU
or
BACK
to save the setting(s) and return to the Sensor Settings menu.
5. Press
MENU
or
BACK
to save the setting and return to the Sensor Settings menu.
Non-O
2
sensor O
2
sensor