AirLink 6940XP Operator’s Manual
Specifications
•
10
Junction Box
Hazardous Location
Certification
ATEX: II 2 G Ex d IIB+H2 Gb;
II 2 D Ex tb IIIC Db IP66
cUL/UL: Class I Groups B, C, and D;
Class II Groups E, F, and G;
Class III
IECEx: Ex D IIB + H2 Gb
Ex tb IIIC Db IP66
Tamb -34°C to +100°C
Sensor Housing
Hazardous Location
Certification
Class I, Division 1 (or Division 2), Groups B, C, D
Ex db IIB Gb
Tamb -40°C to +60°C
NOTE:
Certification only applies to some gases. Toxic
gases not certified for Div. 1 locations are suitable for
Div. 2 applications but have no third party approval.
Battery Type
Lithium-thionyl (Li-SOCl2) battery pack, 3.6 VDC 76 Ah
Battery Life
6 months (45 days for versions with PID)
NOTE:
When the battery indicator is blank, approximately 2 weeks of battery life remain.
Operating
Temperature Range
-40°C to +60°C (-40°F to +140°F)
Humidity Range
0 - 98% relative humidity, non-condensing
Radio Output
• 900 MHz - 52 networks, 255 sensors per network
OR
• 2.4 GHz - 78 networks, 255 sensors per network
RF Connection
N-female radio frequency (RF) connector
Display
Transreflective (sunlight-readable)
160 x 104 LCD screen
LED back-light
Max Cable Length for
Remote-Mounted
Antenna
100 feet
Enclosure Material
Aluminum
Sensor Housing
Material
303 Stainless steel
Max Cable Length for
Remote-Mounted
Sensor Kit
• Electrochemical (EC): 250 feet
• Infrared (IR): 40 feet
• PID: 35 feet
Dimensions
7” D x 11” W x 24” H
Weight
25 lbs.