T3A Transmitter • 61
Register Address B/11: Fault Code
Register Address C/12: Sensor Type
Register Address 17/23: Relay Setting
Register Address 19/25: Calibration Type
Response
Fault Type
0
No Fault
1
Loss of Communication with
Sensor Board
4
Loss of Communication with
Sensor Element/Housing
5
Zero Error
6
Calibration Error
Response
Sensor Type
0
EC – Electrochemical
1
IR – Infrared
2
CB – Catalytic Bead
4
PID – Photo Ionization Detector
Bit
Relay Setting
Function
5
Relay 2: Increasing/
Decreasing Setting
0 – Decreasing
1 - Increasing
4
Relay 1: Increasing/
Decreasing Setting
0 – Decreasing
1 - Increasing
3
Relay 2: Failsafe Setting
0 – No (Off)
1 – Yes (On)
2
Relay 1: Failsafe Setting
0 – No (Off)
1 – Yes (On)
1
Relay 2: Latch/Auto
Reset
0 – Auto Reset
1 - Latch
0
Relay 1: Latch/Auto
Reset
0 – Auto Reset
1 – Latch
Response
Calibration Type
0
Manual Calibration
1
Auto Calibration
Содержание 71-0533