36 • T3A Transmitter
2. Press and release the MENU button until the Modbus Address screen appears.
3. Use the ADD and SUB buttons to increase and decrease the Modbus address number,
respectively.
4. Press and release the MENU button to move through the Product Settings and
Configuration menu or to exit to the normal operating mode screen.
Modbus Baud Setting
The baud rate is the speed of data transmitted within the Modbus system, measured in bits per
second (bps). For successful communication, the baud rate setting of the T3A MUST match
the baud rate setting on the connected controller or other Modbus device.
The T3A’s default Modbus baud setting is 9600 bps. The RS-485 Modbus communication
parameters used in the T3A are 8 data bits, no parity bit, and 1 stop bit (8-N-1) these
parameters are fixed and cannot be changed. Some devices come with different Modbus baud
rates. Check with your system administrator to determine if a different Modbus baud setting is
needed for your system.
The pre-set Modbus baud setting available for the T3A are the following:
•
110 bps
•
300 bps
•
1200 bps
•
2400 bps
•
4800 bps
•
9600 bps
•
19200 bps
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
the MENU button for 6 seconds.
2. Press and release the MENU button until the Modbus Baud screen appears.
3. Use the ADD and SUB buttons to scroll through the available Modbus baud options.