T3A Transmitter • 51
Calibrating the Sensor (Auto Cal)
You should ONLY calibrate the sensor after completing the zero process.
1. Press the ADD button to select “Yes” to begin the calibration process and to advance to
the calibration confirmation screen. If you do not wish to calibrate the sensor, press the
SUB button to select “No” to advance to the sensor radio address setting screen.
2. Press the ADD button to select “Yes” to confirm that you want to calibrate the sensor and
to continue to the concentration setting screen. If you do not wish to continue to calibrate
the sensor, press the SUB button to select “No” to advance to the sensor radio address
setting screen.
3. Use the ADD and SUB buttons to adjust the concentration match the concentration shown
on the calibration cylinder.
For ClO
2
sensors, adjust the ClO
2
concentration to match the Cl
2
concentration shown on
the calibration cylinder. There is no conversion factor.
4. Press the MENU button to save the gas concentration setting and to advance to the
calibration start screen.
5. Unscrew and remove the rain guard from the assembly.
6. Install the calibration cup to the T3A’s sensor housing.
7. Screw the regulator into the calibration cylinder.
8. Use the sample tubing to connect the regulator to the calibration cup.