20 • T2A Transmitter
7. Allow the detector to warmup for the appropriate amount of time as shown below,
depending on the sensor type.
8. The T2A is factory-calibrated before shipping from RKI. If a full calibration is desired at
startup, see page 35.
Detection Gas
Warmup Time
After an Extended
Time Off Power
Warmup Time
After a Short
Time Off Power
Ammonia
12 hours
4 hours
Carbon Monoxide
2 hours
10 minutes
Chlorine
Chlorine Dioxide
Hydrogen Chloride
12 hours
Hydrogen Cyanide
Hydrogen Fluoride
2 hours
Hydrogen Sulfide
Nitric Oxide
12 hours
12 hours
Nitrogen Dioxide
2 hours
10 minutes
Oxygen
Ozone
Phosphine
Sulfur Dioxide