30 • T2A Transmitter
3. Press the ADD button to return the device to its factory default settings and to advance to
the return to factory default settings confirmation screen. If you do not wish to return the
device to its factory default settings, press the SUB or MENU button to continue to the
Reset Zero & Cal Only screen.
4. If “Yes” is selected to return the device to its factory default settings:
5. Press the ADD button to select “Yes” to confirm that you want to reset the device to its
factory default settings and to return the device to normal operating mode. If you do not
wish to continue to return the device to its factory default settings, press the SUB button to
select “No” to continue to the Reset Zero & Cal Only screen.
6. Use the MENU button to scroll through the rest of the Product Settings and Configuration
menu and return to Normal Operating Mode.
NOTE:
If the T2A is reset to the factory default settings, ALL configuration steps MUST be
repeated and the device MUST then be zeroed and calibrated for proper operation of
the device.
Reset Zero & Calibration Values
Resetting the zero and calibration settings of the sensor element will allow the currently stored
zero and calibration values to be rest without having to reconfigure all of the other operational
settings like with the Return to Factory Defaults option.
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
the MENU button for 6 seconds.
2. Press and release the MENU button until the Reset Zero & Calibration Values screen
appears.
3. Press the ADD button to select “Yes” to reset the Zero and Calibration values and to
advance to the Reset Zero & Cal Only confirmation screen. If you do not wish to reset the
zero and calibration values, press the SUB button to select “No” to leave the Product
Settings and Configuration menu and to return the device to normal operating mode.