background image

 

 14 • 65-2427-05-01 H

2

S Detector

Replacing Components of the H

2

S Detector

This section includes a procedure to replace the H

2

S sensor and the entire detector assembly. 

In most cases, it is not necessary to replace the entire detector assembly.

Replacing the Plug-In H

2

S Sensor

CAUTION:

The sensor contains electrolyte which is a dilute acid. Do not disassemble the 

sensor when replacing it with a new one. If sensor electrolyte comes in contact 
with your skin, wash affected area thoroughly with soap and water.

1. Turn off the controller.

2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose 

the cap gasket.

4. Unplug and remove the H

2

S sensor with the rubber boot and spacer attached.

5. Remove the rubber boot and spacer from the old sensor.
6. Remove the replacement sensor from its packaging and remove the wire jumper. This wire 

jumper is installed on the sensor pins for shipment or storage but must be removed for the 
sensor to operate properly when installed in a detector.

Slow or No 
Response/
Difficult or 

Unable to 
Calibrate

• Unable to 

accurately set the 

zero or response 

reading during 

calibration.

• Detector requires 

frequent calibration.

Note:

Under 

“normal” 
circumstances, the 
detector requires 
calibration once every 
3 months.

Some applications 

may require a more 
frequent calibration 
schedule.

• The calibration 

cylinder is low, out-

dated, or defective.

• The flame arrestor in 

the detector housing 

cap is wet or clogged 

with dirt or other 

particulates.

• The calibration gas is 

not an appropriate 

concentration.

• The detector is 

malfunctioning.

1. Verify that the calibration cylinder 

contains an adequate supply of a 

fresh test sample.

2. Check the detector housing cap to 

determine if the flame arrestor is wet 

or dirty. Clean if necessary.

3. Verify that the calibration gas 

concentration is appropriate for the 

detector. Zero emission air (20.9% 

oxygen) is normally used for a zero 

adjustment if the environment is 

suspect and 25 ppm H

2

S in nitrogen 

is normally used for a span 

adjustment.

4. If the calibration/response difficulties 

continue, replace the plug-in sensor 

as described later in this section.

5. If the calibration/response difficulties 

continue, contact RKI for further 

instruction.

Condition

Symptom(s)

Probable Causes

Recommended Action

Содержание 65-2427-05-01

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2427 05 01 Hydrogen Sulfide Detector Operator s Manual Part Number 71 0454 Revision 0 Released 6 15 20...

Страница 2: ...calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type...

Страница 3: ...nt costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRE...

Страница 4: ...Wiring the H2S Detector to a Controller 10 Startup 12 Introducing Incoming Power 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 13 Replacing Components of the H...

Страница 5: ...instructions and warnings in this manual to assure proper and safe operation of the 65 2427 05 01 and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2427...

Страница 6: ...section describes the components of the 65 2427 05 01 H2S detector The 65 2427 05 01 consists of the 65 2428 01 H2S detector and junction box Figure 1 H2S Detector Component Location Terminal Block R...

Страница 7: ...sing and cap A flame arrestor guard is permanently bonded to the cap Two wires extend from the top of the detector housing body Use these wires to connect the H2S detector to the junction box terminal...

Страница 8: ...install a calibration cup on the detector during calibration The terminal block within the junction box facilitates the wiring process A cover on the front of the junction box allows access to the int...

Страница 9: ...e Mounting Dimensions 2 At the monitoring site you select hang or mount the junction box with the detector facing down see Figure 3 38 75 3 10 Rubber Spacers 3X 3 4 NPT Female 3 65 1 1 2 20 Thread for...

Страница 10: ...hazardous location rating of the junction box The junction box is rated NEMA 4X and Class 1 Division 1 Groups B C D 5 Connect the two wires to the detector using the 2 point terminal block CAUTION If...

Страница 11: ...ing the H2S Detector to a Controller 9 If using shielded cable connect the cable s drain wire to an available chassis ground at the controller RKI controllers typically have a ground stud that is a co...

Страница 12: ...g cap or junction box cover while the circuits are energized unless the area is determined to be non hazardous Keep the detector housing cap and junction box cover tightly closed during operation CAUT...

Страница 13: ...the H2S detector NOTE This troubleshooting guide describes detector problems only See the controller operator s manual for problems you may encounter with the controller Table 2 Troubleshooting the H2...

Страница 14: ...or No Response Difficult or Unable to Calibrate Unable to accurately set the zero or response reading during calibration Detector requires frequent calibration Note Under normal circumstances the dete...

Страница 15: ...ibration section of this manual Replacing the H2S Detector NOTE In most cases it is only necessary to replace the H2S sensor 1 Turn off the controller 2 Turn off or unplug power to the controller 3 Re...

Страница 16: ...ample if the environment is not well controlled then more frequent calibration than every 3 months may be necessary Calibration This section describes how to calibrate the H2S detector It includes pro...

Страница 17: ...2S gas cylinder If necessary to fit the H2S calibration gas cylinder change the regulator Setting the Span Reading 8 Follow the directions in the controller operator s manual for setting the response...

Страница 18: ...bber spacers 65 2039RK H2S replacement sensor 65 2427 05 01 H2S detector j box CSA classified 65 2428 01 H2S detector with plug in sensor 71 0454 65 2427 05 01 H2S Detector Operator s Manual this docu...

Отзывы: