Overview
This manual describes the 35-3010RKA-04H sample-draw detector. This manual also describes
how to install, start up, maintain, and calibrate the detector. A parts list at the end of this manual
lists replacement parts and accessories for the sample-draw detector.
Specifications
Table 1 lists specifications for the 35-3010RKA-04H.
WARNING: When using the 35-3010RKA-04H, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the
35-3010RKA-04H and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 35-3010RKA-04H as described in this
manual.
Table 1: Specifications
Target Gases & Detection Ranges
Hydrogen (H
2
): 0 - 100% LEL
Oxygen (O
2
): 0 - 25% volume
Input Power
24 VDC
Current Draw
275 mA
Output Signals
4-20 mA each channel
Construction (housing)
Fiberglass/polyester (NEMA 4X)
Dimensions
15.44 in. H x 12.55 in. W x 8.31 in. D
Weight
14 lbs.
Sampling Method
Sample-draw
Sample Flow
1.2 SCFH (nominal)
Response Time
90% in 30 seconds
Accuracy
Hydrogen:
± 5% of reading or ± 2% LEL (whichever is greater)
Oxygen:
± 0.5% O
2