![RKI Instruments 35-3010RKA-04H Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/rki-instruments/35-3010rka-04h/35-3010rka-04h_operators-manual_1466916013.webp)
35-3010RKA-04H Sample-Draw Detector Head
Description
•
13
Oxygen Transmitter
The oxygen transmitter is mounted to the right of the LEL transmitter and above the main circuit
board. The amplifier includes the amplifier type selector, two internally wired terminal strips,
span pot, zero pot, and test points.
Figure 4: Oxygen Transmitter
Transmitter Type Selector
The transmitter type selector is near the bottom left corner of the amplifier. It is to the left of the
detector terminal strip and below the span pot.
The transmitter included with the sample-draw detector is designed for use with RKI’s oxygen
and toxic gas sensors. The transmitter type selector determines for which sensor the amplifier is
intended. For oxygen transmitters, a jumper block is installed over the
OXY
selector.
Detector Terminal Strip
The detector terminal strip is the four-point terminal strip near the bottom edge of the transmitter.
It is factory wired to the transmitter.
Interconnect Terminal Strip
The interconnect terminal strip is the four-point terminal strip near the top edge of the transmitter.
It is factory wired to the main circuit board.
Span/Zero Pots
The span and zero pots are located on the left side of the transmitter and are used for calibration.
Use the span pot to make adjustments to gas response readings and the zero pot to make
adjustments to the zero reading.
Test Points
The test points are located on the top of the transmitter on either side of the transmitter
interconnect terminal strip. They are labeled
TP+
and
TP-
. A 100 mV - 500 mV output is
available at these test points for use during calibration.
FB
Oxygen
Transmitter
24V
4/20
T
P
-
T
P
+
BATT
+
OXY
+
OXY
W
BK
ZERO
RD
Test Point TP+
Transmitter Type Selector,
Set To OXY
Interconnect
Terminal Strip
Span Pot
Zero Pot
Sensor Terminal
Strip
Test Point TP-
TOXIC
SPAN
G
TOXIC