7. TROUBLESHOOTING 

 

Problem Possible 

cause 

Indication pointer 
completely defects 
to the left or right. 

 

 No indicator input circuit connected. 

 No 4-conductor shielded cable connected. 

 The defective connector used (standard or water 

resistant connector). 

 Wires disconnected or shorted. 

 No internal sensor wiring connected. 

 The fiberglass coated cable immersed into water or 

exposed to high temperature, resulting in 
deteriorated insulation resistance. 

 No rated output set to the PCT-300 (different gain). 

 The double gain selector switch turned ON. 

 The strain gauge deteriorated due to exposure to 

high temperature. 

 No zero adjuster adjusted.

 

 

Digital display 
overflows or 
underflows. 

No pressure is 
indicated under 
pressurized 
condition. 

 Irregularly tapped hole for installing the CZ-200P. 

(The sensor tip strongly contacting with the tapped 
hole.) 

 The diaphragm deteriorated, deformed or 

damaged.* 

 Mechanical Lead pipe deformation by external 

force. 

Pointer or indication 
fluctuates during 
relay actuation. 

 No measures for relay spark killing taken. 

 No 4-conductor cable shield perfectly wired or 

grounded. 

 The PCT-300 located near magnetically operated 

relays. 

Pressure indication 
is fluctuated. 

 Value different from the sensor rated output set 

to the rated output setter for the PCT-300. 

 The diaphragm deteriorated, deformed or 

damaged.* 

 The sensor exposed to hot or cold wind. 

 Some potential against the earth generated 

(Two-point grounding, etc.). 

Normal operation 
was performed, but 
no reading was 
received after a 
while or the reading 
varied and was 
unstable.

 

 Imperfect connecter contact. 

 The lead pipe deformed by external force. 

 The diaphragm deteriorated, deformed or   

damaged.*

 

 The fiberglass coated cable immersed into water or 

exposed to high temperature, resulting in 
deteriorated insulation resistance. 

 The sensor exposed to hot or cold wind. 

 The extruders now in unstable operation or 

temperature rise. 

Indication fluctuates 
from the beginning. 

 The sensor tip forcibly tightened due to the small 

tapped hole. 

 The lead pipe cover contacting with the barrel, etc. 

Resin leakage.

 

 The sealed surface deformed or scratched. 

 Foreign material (carbide, etc.) attached on the 

sealed surface. 

 Low sealed surface accuracy   

(parallelism, axis, etc.). 

 No screw threaded down to the extreme end. 

 Tightened at less than appropriate torque or not 

tightened. 

No threads 
regularly engaged 
(no screw removed).

 

 No screw threaded down to the extreme end. 

 Not threaded as conforming to the standard. 

 The screw with burrs used. 

 Tightened at excessive torque. 

 Tightened at temperature different from the initial 

tightening temperature. 

 Foreign material attached on the threaded section, 

or stained. 

Indication of 
thermocouple is 
lower value.

 

 A protecting tube of thermocouple compensating 

leads and thermocouple connector are damaged, 
deformed, or broken down. 

 

For taking measures, also refer to 

3. HANDLING OF THE 

PRESSURE SENSOR

 on page 2. The converter is described on a 

basis of the PCT-300. The operation of the PG500 and REX-PG410 
may differ from that of the PCT-300 described here. 

 

For the cause of diaphragm deterioration, deformation or 
damage, refer to the 

* Main causes

 on page 9. 

 

 

* Main causes:

 

 These causes may arise independently or in mutual relations. 

 Generative cause in operation 

Overpressure

Load pressure exceeding its limit applied. 

Irregular thread
engagement 

The metal pressure sensing surface mechanically scratched 
or chipped off. 

Metal fatigue 

The metal sensing surface fatigued by the application of 
changing or repeating pressure. 

Corrosion 

The pressure sensing surface corroded due to its contact 
with corroding material. 

Abrasion 

The pressure sensing surface worn away due to the mixture 
of fillers, etc. 

Shrinkage 

The pressure sensing surface deformed due to the shrinkage 
of resin adhered to its surface as a result of extruder cooling 
down. 

Separation 

The pressure sensing surface deformed or damaged by resin 
adhered to its surface due to the removal of the sensor while 
the resin is not yet melted or its melted condition is imperfect.

Protrusion 

The pressure sensing surface deformed or damaged due to 
the protrusion of the push rod from the lead pipe as a result 
of an external force applied by the screen changer, etc. 

Contact 

The pressure sensing surface deformed due to the forced 
contact of the sensor outer side with the hole inner surface 
as a result of the finish of the hole inner surface. 

 

 Generative cause in mounting and removing 

Impact 

The pressure sensing surface deformed due to its strong 
strike with solid material. 

Dropping 

The pressure sensing surface or its circumference scratched 
or deformed due to sensor dropping. 

Handling 

After the resin attached to the sensor tip is heated by a gas 
burner, etc. for its removal at the time of inspection, the 
pressure sensing surface scratched with a metal brush, etc. 

Excessive 
tightening 

The diaphragm deformed or damaged as the push rod pushes 
the diaphragm from the inside as a result of the deformation 
of the sensor flange by excessive tightening torque. 

 

Diaphragm

Flange (Packing/Sealing surface)   
[Flat tightening or tapered tightening surface] 

Sensor pressure sensing part 

 

 

8. SPECIFICATIONS 

„

 

Pressure

 

measurement

 

section 

 

Specification

 

 

Sensing block construction:

 

4-side adhesion-type strain gauge, Wheatstone bridge 

 

Rated pressure: 

Fixed nut type:  10 MPa, 20 MPa, 35MPa, 50 MPa, 70 MPa, 100 MPa, 
  

150 

MPa 

Loose nut type:  0.5 MPa *, 1 MPa, 2 MPa, 3 MPa, 5 MPa, 10 MPa, 
   

20 MPa, 35MPa, 50 MPa, 70 MPa, 100 MPa 

 

*  However, this range can be used when combined with the   

 

  CT-300 (the ZK-872 specification), PG500 or REX-PG410 

 

The SUPRON diaphragm specification type is not available to the 
range of 5 MPa or less and 150 MPa. 

 

Rated output: 

1.0 to 1.8 mV/V 

  [Calibration temperature: At diaphragm temperature of 150 

°

C] 

0.5 to 0.9 mV/V (For the “0.5 MPa” range) 

  [Calibration temperature: At diaphragm temperature of 150 

°

C] 

 

The output of each sensor becomes a specific value within the 
range of 1.0 to 1.8 mV/V. 

 

Bridge impressed voltage:

 

  10 V DC (When using PCT-300 or CT-300),   
  7.7 V DC (When using PG500 or REX-PG410) 

Accuracy: 

SUS630 diaphragm specification type 

[At diaphragm temperature of 150 

°

C]

:

 

  The range of 70 MPa or less: 

±

 1.0 % of span 

  The range of 100 MPa or more: 

±

 2.0 % of span 

SUPRON diaphragm specification type 
  The range of 70 MPa or less: 

±

 1.0 % of span 

  More than 480 

°

C of 10 MPa, 20 MPa and 70 MPa range: 

±

 2.0 % of span 

  The range of 100 MPa: 

±

 2.0 % of span 

  

480 

°

C or more: 

±

 4.0 % of span 

 

Содержание CZ-200P

Страница 1: ...rned z Tools such as wire wheels or abrasive cloths should never be used to clean the process diaphragm z Dropping the instrument may cause the sensor to break or malfunction z To protect built in precision components from damage handle and install carefully Take care to avoid damage to the diaphragm z Do not wipe or rub the nameplate on the outer case of the resin pressure sensor with a cloth moi...

Страница 2: ...C diaphragm specification is not available 4 The SUPRON diaphragm specification is not available to the range of 5 MPa or less and 150 MPa z Accessories Instruction Manual IM100CZ08 E9 Copper Packing Included in only the Loose nut type thickness t 2 mm If any of the products are missing damaged or if your manual is incomplete please contact RKC sales office or the agent Temperature sensor cable So...

Страница 3: ...the exposed lead pipe is further extended In this case take measures of 1 The effect of heat flow lessens as the installing direction of the sensor changes from the slanting direction Fig 1 B to the horizontal direction Fig 1 C in this order In this case take measures of 1 and 2 if necessary by checking the sensor outer cylinder surface temperature To the relevant manufacturer It is recommended th...

Страница 4: ...rush or bend the thermocouple protecting flexible tube extended from the outer cylinder or the thermocouple plug If so the thermocouple may be damaged to disable temperature measurement or a temperature error may result The temperature sensing block is located about 2 mm within the diaphragm and as a result the measured temperature usually becomes lower than the actual temperature depending on the...

Страница 5: ...pe cover Optional The outer tube and the cable section of the loose nut type and the unified thread type with directly connected cable are the same as the above figure The mounting hole dimensions are the same as those for the standard type PF3 8 JIS B grade Diaphragm Unit mm External dimensions Standard dimensions H A H B H C H D L 120 150 180 210 L1 8 8 8 8 L2 6 6 6 6 L3 60 90 120 150 Weight App...

Страница 6: ...21 Colors Terminal No Input terminals Wiring with REX PG410 pressure indicator Pressure sensor cable Pressure sensor CZ 200P SHD E EXC A EXC C SIG D SIG B REX PG410 Shield Red Brown Blue Black 6 7 8 9 10 Colors Terminal No Input terminals For details on wiring with PCT 300 PG500 or REX PG410 refer to the relevant instruction manual Wiring example of built in temperature sensor type Optional Pressu...

Страница 7: ...setter Monitor plug Zero adjuster Power lamp PCT 300 PWR ZERO mV V MONI 1 Check the rated output mV V described on the nameplated of the CZ 200P This output should be corrected when the cable is extended and then set that value on the rotary switch which is gain setter of the PCT 300 2 The pressure reading zero point is adjusted by the zero adjuster in PCT 300 Perform this zero adjustment after th...

Страница 8: ...al length L 10 m Rated output of the resin pressure sensor e0 1 500 e1 1 500 1 1 96 10 4 10 5 e1 1 499 Set 1 499 to the gain setting portion of the converter 6 3 Correcting Indication Error when the Cable of Another Company is Used The nominal cross sectional area of our cable conductor is 0 5 mm 2 If using a cable from another company the resin pressure sensor s indication value can be corrected ...

Страница 9: ...ation or damage refer to the Main causes on page 9 Main causes These causes may arise independently or in mutual relations Generative cause in operation Overpressure Load pressure exceeding its limit applied Irregular thread engagement The metal pressure sensing surface mechanically scratched or chipped off Metal fatigue The metal sensing surface fatigued by the application of changing or repeatin...

Страница 10: ... diaphragm specification type As to diaphragm temperature 0 15 0 2 of span 10 C Mechanical characteristics Allowable over pressure 120 of span 0 5 MPa 1000 of span 1 MPa 500 of span Limited over pressure 150 of span 0 5 MPa 2000 of span 1 MPa 1000 of span Fixing screw section material SUS630 Lead pipe cover material SUS304 Fixed nut type only Diaphragm material SUS630 HASTELLOY C SUPRON Recommende...

Отзывы: