Rish Insu 5Dx
23
2.
If, upon pressing the
START/STOP
key
(with nominal test voltage set to
1000V
and load resistance
5M
), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acoustic signal and displays the
screen reported here to the side for 3
seconds. Afterwards, the instrument
returns to the initial screen.
3.
If, upon pressing the
START/STOP
key
(with 1000V < nominal test voltage set
2500V and load resistance
5M
), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acoustic signal and displays the
screen reported here to the side for 3
seconds. Afterwards, the instrument
returns to the initial screen.
4.
If, upon pressing the
START/STOP
key
(with 2500V < nominal test voltage set
5000V and load resistance
5M
), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acoustic signal and displays the
screen reported here to the side for 3
seconds. Afterwards, the instrument
returns to the initial screen.
5.
If, upon pressing the
START/STOP
key or if,
while measuring, an insufficient level of
the supply battery voltage is detected, so
that the success of the test is prejudiced,
measurement does not start or is stopped.
The instrument gives a long acoustic signal
and displays the screen reported here to
the side for 3 seconds. Afterwards, the
instrument returns to the initial screen.